16 Centrifugation
In accordance with EN / IEC 61010-2-020, no persons, hazardous substances or objects may be within a
safety range 300 mm around the centrifuge during a centrifugation run.
After each centrifugation run, while the lid is unlocking, the remaining number of running cycles (centrifugation
runs) will be briefly displayed.
Example:
PROG t/min:sRPM
If the permissible weight difference within the rotor load is exceeded, the centrifugation run is aborted during
the start-up and the following display appears:
PROG t/min:sRPM
A centrifugation run can be aborted at any time by pressing the
STOP
OPEN
key.
During a centrifugation run, centrifugation parameters can be selected and changed. The changed values are
stored at program location #.
With the
RCF
key, you can toggle between the RPM display (RPM) and RCF display ( RCF ) at any time. If
the RCF display (
RCF ) is worked with, the centrifuging radius must be input.
The following is displayed:
PROG t/min:sRPM
Further operation of the centrifuge is then only possible after opening the lid.
Operation errors and malfunctions will be shown (see the chapter on "Malfunctions").
 Plug in the mains switch. Switch position .
 Load the rotor and close the centrifuge lid.
16.1 Centrifugation with preset time
Example:
RPM display (RPM)
Example:
RCF display (
RCF )
RCF
 Press the key to activate the RPM display
(RPM) or the RCF display ( RCF ) as
needed. The LED above the key is
illuminated if the RCF display (
RCF ) is
activated.
PROG t/min:sRPM PROG t/min:sRPM
 Input the desired centrifugation parameters (see the
"Programming" chapter).
PROG t/min:sRPM PROG t/min:sRPM
START
PULSE
 Press the key.
The centrifugation run is started.
During the centrifugation run, the rotor
speed or the resulting RCF value and
remaining time are displayed.
PROG t/min:sRPM PROG t/min:sRPM
Continued on next page