20.5 Backlighting of the display
To save energy, it can be set that, after a centrifugation run, the backlighting of the display switches off after 2
minutes.
With the rotor at a standstill, this can be set as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key. In this case, the settings are not stored.
Press and hold the button for eight
seconds.
PROG t/min:sRPM
Press the key until the following is
displayed:
PROG t/min:sRPM
START
PULSE
Press the key.
PROG t/min:sRPM
PROG t/min:sRPM
Press the key until one of the following is
displayed:
Power save : Automatic shutdown of the
backlighting.
off : Automatic shutdown deactivated.
on : Automatic shutdown activated.
PROG t/min:sRPM PROG t/min:sRPM
With the keys, set off or on.
PROG t/min:sRPM PROG t/min:sRPM
START
PULSE
Press the key to save the setting.
PROG t/min:sRPM
PROG t/min:sRPM
STOP
OPEN
Press the key once to exit the "
Settings"
menu or press twice to exit the "
MACHINE MENU
".
PROG t/min:sRPM
Example: