EN
82/165
Message / fault Cause Remedy
CTRL ERROR 22 – 25.4
CRC ERROR 27, 27.1
COM ERROR 31 – 36
FC ERROR 60,
61.1 – 61.21,
61.64 – 61.142
Error / defect electronics
 Perform a MAINS RESET when the
rotor has been stationary.
TACHO ERR 61.22
FC ERROR 61.23
Speed measurement error
 The device must not be switched off
as long as the "wait" message is
being displayed. After the message
"wait ...!" is no longer displayed,
carry out a MAINS RESET.
FC ERROR 61.153 Error / defect electronics
 Perform a MAINS RESET.
 Check the loading of the rotor, see
chapter "Loading the rotor".
 Repeat the centrifugation run.
VERS. ERR 61.154 Invalid machine version
 Perform a MAINS RESET.