Specifications
Affinity® 4 Birthing Bed Instructions for Use (USR025 REV 13) 53
IMMUNITY to Proximity Fields from Radio Frequency Wireless Communications Equipment
In addition to the Radiated RF IEC 61000-4-3 as shown in the table above, the P3700E Affinity® 4 Birthing Bed with
serial numbers after T242AA hav been tested as specified in the table below.
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation Maximum
Power (W)
Distance (m) Immunity Test
Level (V/m)
385 380-390 TETRA 400 Pulse
modulation
18 Hz
1,8 0,3 27
450 430-470 GMRS 460,
FRS460
FM +
5 kHz
deviation
1 kHz sine
20,3 28
710
704-787 LTE Band 13, 17
Pulse
modulation
217 Hz
0,2 0,3 9745
780
810
800-960
GSM 800/900
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
20,3 28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,3, 4,
25; UMTS
Pulse
modulation
217 Hz
20,3 28
1845
1970
2450 2400-2570 Bluetooth
WLAN, 802.11
b/g/n,
RFID
Pulse
modulation
217 Hz
20,3 28
5240
5100-5800
WLAN
802.11
a/n
Pulse
modulation
217 Hz
0,2 0,3 95500
5785