Connections (Instrument and Control Device)
(3) Output signals
Signal Description For more information
EOM This signal is output when measurement completes. The
comparator judgment results, ERR signal, and BIN signal are
updated when the EOM signal is output.
“EOM signal output
type” (p. 137)
ERR This signal is output when a contact error (display: NoCntct),
temperature compensation error (display: Err.TC), or other error
occurs.
All comparator judgment result output turns off while the ERR
signal is output. The ERR signal is also output when the instrument
encounters an internal circuitry error or a calculation results error.
“3.5 Measurement
Error Displays
(Displays Other Than
Measured Values)”
(p. 46)
HI,IN,LO
These signals are used to output comparator judgment results.
OB,
BIN0toBIN9
The BIN judgment results are output from pins 13 to 17 and pins
31 to 36 when BIN measurement is set to ON. If the results do not
correspond to BIN0 to BIN9, the OB signal (pin 31) will turn on.
• “4.3 BIN
Measurement (Using
Multiple Judgment
Standards)” (p. 57)
• See explanation on
following page.
OUT0to
OUT10
Pins 13 to 17 and pins 31 to 36 can be used as general-purpose
output pins while BIN measurement is set to OFF. Output signals
can be controlled with the
:IO:OUTPut
command.
• “4.3 BIN
Measurement (Using
Multiple Judgment
Standards)” (p. 57)
• See explanation on
following page.
• Communication
Command Manual
on the included
application disc.
0
The TRIG signal is ignored while the measurement conditions are being changed.