58
BIN Measurement (Using Multiple Judgment Standards)
(Measurement screen) MENU > MATH > BIN
1
Enable the BIN function.
(Default setting: OFF)
When this function is set to OFF, upper limit
and lower limit values are disabled even if
they have been previously set.
2
Select a BIN number.
3
1
2
Enter an upper limit value and touch
ENTER.
CLR Deletes the entire value.
BS Deletes the previous character.
CANCEL Cancels the setting and returns to
the previous screen.
(Default setting: 0 V; valid setting range:
-1000 V to 1000 V)
If the instrument is turned off before touching
ENTER, the value being set will be lost, and
the setting will revert to its previous value.
4
Enter a lower limit value using the same
method as for the upper limit value.
(Default setting: 0 V; valid setting range:
-1000 V to 1000 V)
w ww . . co m
information@itm.com1.800.561.8187