59
BIN Measurement (Using Multiple Judgment Standards)
5
1
2
Enable the BIN number setting.
When the BIN number is set to OFF, upper
limit and lower limit values are disabled even
if they have been previously set.
Judgment result
Set BIN number
Green: Judgment complete
White: Judgment in progress
(BIN numbers set to OFF are not displayed.
The judgment result and sub-display are
shown on the Measurement screen.
Setting the BIN function to ON automatically
sets the comparator function to OFF.
Thefollowingjudgmentsmaybeshownwhentheinstrumentisnotabletoperformmeasurement
normally:
Measuredvaluedisplay BIN judgment
+OvrRng OB (out of range)
−OvrRng OB (out of range)
NoCntct or −−−−−−− −− (No judgment)
See “3.5 Measurement Error Displays (Displays Other Than Measured Values)” (p. 46)
.
TheSettingsscreencanbedisplayedfromtheMeasurementscreen’ssub-display.
Tocheckjudgmentsaurally
Judgment tones do not sound during BIN measurement.
Tooutputjudgmentresultstoanexternaldeviceortoprintjudgment
results
Set the BIN function to ON, congure external output (p. 111) or printing (p. 143), and prepare the
associated equipment.