72
Contact Check
Threshold
The instrument’s contact check threshold is specied as a capacitance value. It can be changed
within the range of 0.5 nF to 50 nF (default setting: 1 nF).
The following table provides approximate resistance value threshold equivalents for a number of
contact check thresholds:
Threshold value
setting
Resistance value
threshold
0.5 nF 15 k
Ω
5 nF 1.5 k
Ω
50 nF 150
Ω
If the capacitance between the HIGH and LOW terminals is less than the threshold value, the
instrument will not display a measured value or perform judgment (contact error). Set a low
threshold for small batteries and a high threshold for large batteries.
The capacitance between the HIGH and LOW terminals can be monitored and used as a
guideline when determining the threshold.
If the capacitance monitor value is the same
as the contact check threshold, the instrument
may either detect a contact error or display the
measured value
.
Contactcheckintegrationtime
The contact check integration time can be changed within the range of 1 ms to 100 ms (default
setting: 10 ms). Use a low value when you wish to increase the measurement speed. Use a high
value in environments with a large amount of noise.
It is recommended to enable the contact check function when switching among multiple
measurement targets at high speed as part of the measurement process and when measuring
the potential across a battery’s enclosure.
Set an appropriate trigger delay (p. 39) when measuring the potential across a battery’s
enclosure. In particular, discharge time is required when the enclosure has been charged.