73
Contact Check
(Measurement screen) MENU > MEAS > CNT.CHECK
1
Enablethecontactcheckfunction.
(Default setting: OFF)
2
1
2
3
Set the threshold (capacitance).
CLR Reverts the setting to its
default value.
CANCEL Cancels the setting and
returns to the previous screen.
(Default setting: 1 nF; valid setting range:
0.5 nF to 50 nF)
3
1
2
3
Setthecontactcheckintegrationtime.
CLR Reverts the setting to its
default value.
CANCEL Cancels the setting and
returns to the previous screen.
(Default setting: 10 ms)