Operating Precautions
10
Before Connecting Data Cables (USB, RS-232C)
Observe the following precautions when connecting the instrument and a
controller:
To avoid faults, do not disconnect or reconnect the USB cable during instru-
ment operation.
The USB and RS-232C interfaces are not isolated from the ground circuit.
Connect the instrument and the controller to a common earth ground.
Using different grounds could result in potential difference between the
instrument and the controller. Potential difference on the data cable can
result in malfunctions and faults.
Before connecting or disconnecting the RS-232C Cable, always turn off the
instrument and the controller. Failure to do so could result in equipment
malfunction or damage.
After connecting the RS-232C Cable, tighten the screws on the connector
securely. Failure to secure the connector could result in equipment mal-
function or damage.
Before Connecting the Printer
Because electric shock and instrument damage hazards are present,
always follow the steps below when connecting the printer.
Always turn off the instrument and the printer before connecting.
A serious hazard can occur if a wire becomes dislocated and contacts
another conductor during operation. Make certain connections are
secure.