GasAlertMicro 5/PID/IR
User Manual
30
Automatic Oxygen (O
2
) Calibration Logger Option
The Logger option is used to define how often the detector records a
datalog sample (once every 1 to 127 seconds).
From the user options menu, press H to scroll to Logger. Press C
within 20 seconds to confirm.
Press G or H to change the current logger rate. When the desired
value displays, press C within 5 seconds to confirm the new value.
When the Autocal option is enabled, it
forces the detector to automatically
calibrate the oxygen sensor during
startup.
If the Autocal option is enabled, ensure
the detector is activated in a safe area
that is free of hazardous gas in an
atmosphere of 20.9% oxygen.
From the Sensor menu, press H to scroll
to O
2
and press C within 20 seconds to
confirm.
Press H to scroll to Autocal. Press C
to toggle between enable/disable.
The detector is shipped with the Autocal
option enabled.