GasAlertMicro 5/PID/IR
User Manual
34
Initialize
A corresponding list of toxic sensors
displays. A checkbox displays beside
the current toxic sensor.
Note
Toxic 1: List includes the PID and CO
2
sensors.
Toxic 2: List includes the H
2
S/CO COSH
sensor.
3. Press G or H to scroll to the new
sensor and press C to confirm. A
checkbox displays beside the new
sensor. To reconfigure, exit the user
options menu.
The following screen displays. The
detector deactivates and immediately
reactivates. It performs the
reconfiguration during the startup.
The new sensor must be calibrated.
Until the new sensor is calibrated, the
due date displays as OL (over limit)
when in normal operation.
4. Calibrate the new sensor
immediately.
The Initialize option restores the origi-
nal factory default settings of the detec-
tor.
1. Press H to scroll to Initialize and
press C within 20 seconds to
confirm.
2. From the Initialize? screen
perform one of the following options
within 5 seconds.
Press A No to exit, or
Press C Yes to initialize.
If No is selected, the following screen
displays and the detector exits the initial-
ize option.