GasAlertMicro 5/PID/IR
User Manual
34
Initialize
A corresponding list of toxic sensors 
displays. A checkbox displays beside 
the current toxic sensor.
Note
Toxic 1: List includes the PID and CO
2
 
sensors.
Toxic 2: List includes the H
2
S/CO COSH 
sensor.
3. Press G or H to scroll to the new 
sensor and press C to confirm. A 
checkbox displays beside the new 
sensor. To reconfigure, exit the user 
options menu.
The following screen displays. The 
detector deactivates and immediately 
reactivates. It performs the 
reconfiguration during the startup.
The new sensor must be calibrated. 
Until the new sensor is calibrated, the 
due date displays as OL (over limit) 
when in normal operation.
4. Calibrate the new sensor 
immediately.
The Initialize option restores the origi-
nal factory default settings of the detec-
tor.
1. Press H to scroll to Initialize and 
press C within 20 seconds to 
confirm.
2. From the Initialize? screen 
perform one of the following options 
within 5 seconds.
• Press A No to exit, or 
• Press C Yes to initialize.
If No is selected, the following screen 
displays and the detector exits the initial-
ize option.