15
GasAlertMicro 5/PID/IR
Activating the Detector
Pump Test
8. If the pump module is attached to the detector, the following
screens display.
When the following screen displays, block the pump inlet.
If the pump inlet is not blocked within 10 seconds or the pump
test fails, the following screens display.
If C is not pressed or the pump is not removed within
25 seconds, the detector performs the pump test again.
Note
The pump may require a new filter. Refer to Replacing the
Pump Filter.
If the pump alarm persists, remove the battery pack to deacti-
vate the detector and then refer to the “Pump Operation” sec-
tion of Troubleshooting
.
If the pump test is successful, the following screen
displays and the self-test continues.