Alarm Setpoints
8. When all of the sensor due dates have been set or bypassed, 
the alarm setpoints need to be set or bypassed. The following 
screen displays.
Factory alarm setpoints may vary by region. Refer to Resetting 
Gas Alarm Setpoints for an example. 
Alarms can be set to any value within the detection range of 
the selected sensor. Refer to “Detection Range” in Specifica-
tions.
Note
           To disable an alarm setpoint, set it to 0 (zero).
If a new setpoint is entered but not confirmed within 5 seconds 
by pressing C, the following screen displays.
The previous setpoint is retained and the detector 
proceeds to the next setpoint.
The setpoints are set in the following order:
• TWA (if applicable)
• STEL (if applicable)
•low
•high
To bypass a setpoint, press C to save the current value and 
proceed to the next setpoint.