Datalogger
Detectors equipped with the datalogger option record information that
can be compiled to create a report. To set how often the detector records
a sample (1-127 seconds), refer to Logger Option
in the user options
menu.
The following information is recorded in a datalog:
Date and time
Serial number of the detector
Type of gas the detector monitors
Gas reading(s) that display
STEL and TWA readings
Sensor status
Detector status
Passcode protect enabled/disabled
STEL period setting
Confidence beep enabled/disabled
Automatic backlight enabled/disabled
Stealth mode is enabled/disabled
Latching alarm enabled/disabled
Calibration past due option enabled/disabled
Language the detector is set to display
MMC/SD Card Compatibility
a Caution
To ensure the Intrinsic Safety rating of the detector, use
only the following certified MultiMediaCards (MMC) and
Secure Digital (SD) cards.
The following MMCs and SD cards are approved for use with the
GasAlertMicro 5 and GasAlertMicro 5 PID detectors:.
128 MB Delkin SD card
64 MB Unigen SD card
The following MMCs and SD cards are approved for use with the
GasAlertMicro 5 and GasAlertMicro 5 PID detectors:.
128 MB Delkin MMC
128 MB Delkin SD card
128 MB Transcend MMC
64 MB Unigen SD card
To purchase additional MMC or SD cards, refer to Specifications
.
Inserting the MMC/SD Card
To insert the MMC/SD card into the detector, refer to the following steps
1-5 and Figure 12.
.
1. Deactivate the detector.
2. Release the latch and remove the battery pack.
3. Insert the MMC/SD card (pins face down).
4. Replace the battery pack and secure the latch.