GasAlertMicro 5/PID/IR
User Manual
56
Setting the Remaining Alarm Setpoints
9. Repeat step #8 to set the alarm setpoints for the remaining 
sensors. When complete, the detector emits two quick beeps 
and proceeds to Finish Calibration
.
Finish Calibration
10. The detector displays the following screen to indicate that 
calibration is complete.
Verification
1. After calibration is complete and the detector returns 
to normal operation, verify the calibration by using a 
gas cylinder other than the one used for calibration. 
2. The gas concentration should not exceed the sensor's detec-
tion range. Confirm that the LCD displays the expected con-
centration values.
3. To ensure the reading is accurate, apply the verification gas for 
the same amount of time as was applied to the sensor when it 
was calibrated.
Example: SO
2
 span time 2 minutes therefore, apply 
verification gas for 2 minutes.