GasAlertMicro 5/PID/IR
User Options Menu
In the following order, press and continue to hold each button until
Tech mode displays below the Language option.
1. Press and hold H with the right index finger.
2. Press and hold G with the right middle finger.
3. Press and hold C with the left thumb.
4. Press C to enter Tech mode. The options are as
follows:
Sensors
Initialize
Force Calibration (Force calibration)
Bump Daily
(Bump test daily)
Stealth Mode
Sleep Mode
Sensors
a Caution
To reconfigure the sensor type, physically change the
sensor prior to entering Tech mode.
When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
Note
If a sensor is replaced, the detector will classify the sensor as
overdue for calibration. Calibrate the new sensor immediately.
1. Press H to scroll to Sensors. Press
C within 20 seconds to confirm and
access the toxic sensor menu.
2. Press G or H to scroll to Toxic 1
or Toxic 2 and press C within
20 seconds to confirm.