GasAlertMicro 5/PID/IR
Troubleshooting
Troubleshooting
If a problem occurs, refer to the solutions provided in Table 21.. If the problem persists, contact BW Technologies by Honeywell.
Table 21. Troubleshooting
Problem Possible Cause Solution
Startup Troubleshooting
The detector does not
activate.
No batteries Refer to Replacing the Alkaline Batteries
or Replacing the Lithium Battery
Pack.
Depleted batteries Refer to Replacing the Alkaline Batteries
, or Replacing the Lithium Battery
Pack.
Damaged or defective detector Contact BW Technologies by Honeywell
.
The detector immediately
enters alarm mode when
activated.
Sensor needs to stabilize Used sensor: wait 60 seconds.
New sensor: wait 5 minutes.
Low battery alarm Refer to Battery Cautions
, Replacing the Alkaline Batteries, Replacing the
Lithium Battery Pack.
Sensor alarm Refer to Replacing a Sensor or Sensor Filter
.
Pump alarm If the sampling hose is attached, determine if it is obstructed. If not, clean
or replace the pump filter. If the pump alarm persists, refer to the “Pump
Operation” section in Troubleshooting.
The startup self-test fails.
General fault Ensure that the sensors and battery pack are installed correctly and then
reactivate the detector.
If the fault persists, record the error message and contact BW Technologies
by Honeywell.