5
Table 5-3. Command List
Command Parameter
Response
Unit Min Max Default Notes
:DISPlay
:ENABle OFF
j
ON
j
0
j
1 1
:ENABle? 0
j
1
:LOCK OFF
j
ON
j
0
j
1,
<
value
>
password
:LOCK? 0
j
1
:OUTPut
[
:STATe
]
OFF
j
ON
j
0
j
1 0
[
:STATe
]
? 0
j
1
[
:SOURce
]
:AM
:INTernal
:FREQuency
<
value
>
j
MIN
j
DEF
j
MAX HZ
250 300 000 80 000
:FREQuency?
<
value
>
HZ
MIN
j
DEF
j
MAX
<
value
>
HZ
:SOURce INT
j
INT1
j
INT2
j
EXT
j
0
j
1
j
2 0
:SOURce? 0
j
1
j
2
:STATe OFF
j
ON
j
0
j
1 0
:STATe? 0
j
1
[
:SOURce
]
:MODout FRQ
j
FRQ&RDY
j
0
j
1
[
:SOURce
]
:MODout? 0
j
1
[
:SOURce
]
:POWer
:ATTenuation
<
value
>
j
MIN
j
DEF
j
MAX
[
DB
] 40.0dBm
y
0.0dBm
y
0.0dBm
:ATTenuation?
<
value
>
DB
MIN
j
DEF
j
MAX
<
value
>
DB
:AUTO OFF
j
ON
j
0
j
1 0
:AUTO? 0
j
1
:DARK OFF
j
ON
j
0
j
1 0
:DARK? 0
j
1
[
:LEVel
]
[
:IMMediate
]
[
:AMPlitude
]
<
value
>
j
MIN
j
DEF
j
MAX DBM
-10dBm
y
-4dBm
y
-7.0dBm
W
100
W
y
398
W
y
200
W
Attenuator Option
DBM
-50dBm
y
-5.5dBm
y
-7.0dBm
W
10nW
y
300
W
y
200
W
[
:AMPlitude
]
?
<
value
>
DBM
,
W
MIN
j
DEF
j
MAX
<
value
>
DBM
,
W
:UNIT DBM
j
DBMW
j
W DBM
:UNIT? DBM
j
W
Remote Commands 5-3