Execution Errors
These are error messages in the range -200 to -299. They indicate that an
execution error has been detected by the execution control block.
An execution error is signaled by the execution error bit (bit 4) in the event
status register.
-200 Execution error.
This indicates that an execution error has occurred but
the control block cannot be more specic
.
-201 Invalid while in local.
This command is invalid because it conicts with
the conguration under local control.
-202 Settings lost due to rtl.
A local setting was lost when the instrument was
changing from remote to local control, or from local to remote control.
-220 Parameter error.
This indicates that a parameter error has occurred but
the control block cannot be more specic.
-221 Settings conict.
A valid parameter was received, but could not be used
during execution because of a conict with the current state of the instrument.
-222 Data out of range.
The data, though valid, was outside the range allowed
by the instrument.
-223 Too much data.
The block, expression, or string data was too long for the
instrument to handle.
-224 Illegal parameter value.
One value from a list of possible values was
expected. The parameter received was not found in the list.
-240 Hardware error.
Indicates that a command could not be executed due to
a hardware error but the control block cannot be more specic.
-241 Hardware missing.
Indicates that a command could not be executed
because of missing instrument hardware.