5
:TRACe:POINts?
<
trace name
>
Syntax
:TRACe:POINts?
Description
This query returns the number of data points that are in the
trace. Each data point is described by an x- and a y-value.
Example
OUTPUT 724;":TRAC:POIN? CC_UNCAL"
ENTER 724;A$
:TRACe[:DATa]?
<
trace
name
>
Syntax
:TRACe
[
:DATa
]
?
Description
This query returns the trace data for the named trace
. The
information is returned as a string with the values separated by
commas (
,
). The values are grouped into data points. Each data
point has an x-value and a y-value. The x-values are given in
meters, the y-values are given in Watts.
Example
OUTPUT 724;":TRAC? CC_UNCAL"
ENTER 724;A$
Other Commands
WAVEACT
Syntax
WAVEACT
<
wsp
><
value
>
where
<
value
>
is the wavelength. If no units are specied,
<
value
>
is taken to be in meters.
Description
This command allows the Tunable Laser Source to adjust the
wavelength at least
6
1pm of the wavelength meter reading.
Note
The accuracy of WAVEACT is dependent on the wavemeter
specication.
It is only as exact as the Wavelength meter.
5-38 Remote Commands