5
[:SOURce]:AM:INTernal:FREQuency?
Syntax
[
:SOURce
]
:AM:INTernal:FREQuency?
[
MIN
j
DEF
j
MAX
]
Description
This query returns the frequency of the amplitude modulation,
as a îoating point number in Hertz. If you specify
MIN
,
DEF
,or
MAX
with the query, it will return, respectively, the minimum,
default, or maximum modulation frequency.
Example
OUTPUT 724;":AM:INT:FREQ 40.4KHZ"
OUTPUT 724;":AM:STAT ON"
OUTPUT 724;":AM:INT:FREQ?"
ENTER 724;A$
[:SOURce]:AM:SOURce
Syntax
[
:SOURce
]
:AM:SOURce
<
wsp
>
INT
j
INT1
j
INT2
j
EXT
j
0
j
1
j
2
Description
This command chooses the type or source of the modulation of
the laser output.
Set the source to
INT
,
INT1
or
0
to select internal modulation,
set the source to
INT2
or
1
to select coherence control, or
set the source to
EXT
or
2
to select external modulation. The
default is for internal modulation to be selected.
Modulation
Input
When external modulation is selected, the signal is modulated
according to the signal applied to the modulation input on the
front panel of the instrument. The maximum signal that can be
applied is 5V
pp
, which results in 15% modulation of the optical
output.
[:SOURce]:AM:SOURce?
Syntax
[
:SOURce
]
:AM:SOURce?
Description
The query returns the current state of modulation.
A returned value of
0
indicates that internal modulation is
selected. A returned value of
1
indicates that coherence control
is selected,
2
indicates that external modulation is selected.
Example
OUTPUT 724;":AM:SOUR ON"
OUTPUT 724;":AM:SOUR?"
ENTER 724;A$
Remote Commands 5-19