5
:OUTPut[:STATe]?
Syntax
:OUTPut
[
:STATe
]
?
Description
The query returns the current state of the laser current.
A returned value of
0
indicates that the laser current is o. A
returned value of
1
indicates that the laser current is on.
Example
OUTPUT 724;":OUTP ON"
OUTPUT 724;":OUTP?"
ENTER 724;A$
[:SOURce] Commands
[:SOURce]:AM:INTernal:FREQuency
Syntax
[
:SOURce
]
:AM:INTernal:FREQuency
<
wsp
>
<
value
>
j
MIN
j
DEF
j
MAX
Description
This command sets the frequency of the amplitude modulation
of the laser output.
The frequency can be set in the range 250Hz to 300kHz. You
can set to the minimum, default or maximum programmable
values by sending
MIN
,
DEF
,or
MAX
, respectively, instead of the
value.
The default units are
HZ
, though
KHZ
,
MAHZ
,
GHZ
,
THZ
can
alternatively be specied.
The resolution of the frequency is given in the table below:
Range (Hz) Resolution (Hz)
250 ... 999 1
1 000 ... 9990 10
10 000 . . . 99 900 100
100 000 . . . 300 000 1000