5
Example
OUTPUT 724;":WAVE:REF:STAT OFF"
OUTPUT 724;":WAVE 1542E-9"
OUTPUT 724;":WAVE?"
ENTER 724;A$
[:SOURce]:WAVElength:REFerence?
Syntax
[
:SOURce
]
:WAVElength:REFerence?
Description
The query returns the reference wavelength (
0
) in meters.
Example
OUTPUT 724;":WAVE:REF?"
ENTER 724;A$
[:SOURce]:WAVElength:REFerence:DISPlay
Syntax
[
:SOURce
]
:WAVElength:REFerence:DISPlay
Description
This command sets the reference wavelength to the value of the
output wavelength (
!
0
).
Example
OUTPUT 724;":WAVE:REF:DISP"
[:SOURce]:WAVElength:FREQuency
Syntax
[
:SOURce
]
:WAVElength:FREQuency
<
wsp
><
value
>
Description
This command sets the frequency dierence used to calculate
a relative wavelength. The output wavelength is made up
of the reference wavelength and this frequency dierence.
The default units for the frequency are Hertz. The output
wavelength (
) is set from the base wavelength (
0
) and the
frequency oset (df). The formula for calculating the output
wavelength is:
=
c
(
0
)
df
+
c
0
where c is the speed of light in a vacuum (2.998
2
10
8
ms
-1
).
[:SOURce]:WAVElength:FREQuency?
Syntax
[
:SOURce
]
:WAVElength:FREQuency?
5-28 Remote Commands