604
Description
Cause
What
To
Do
605
Description
Cause
What To Do
608
Description
Cause
What To
Do
611
Description
Cause
What
To
Do
6
12
Description
Cause
What
To
Do
+38Vpouer supply failure
The
+38
V
diagnostic test point decreased by more
than approximately
2.2
V.
The
+38
V
supply on the
A2
power supply has failed.
Check the
+38
V
supply.
ALC
out-of -lock div het main band
The
ALC
loop
is
out of lock in the indicated bands.
Either the
A6
output has failed, or the signal from
the
A5
synthesizer has failed or the
A4
reference has
failed.
If
only the het band has failed it is possible
that the
A4
reference
has
failed.
Check the input to the
A6
output from the
A5
synthesizer and the
A4
reference.
Synth out-of-lock divhetmainband
The synthesis loop is out of lock in the indicated
bands.
Either the
A5
synthesizer has failed, or the signal
from the
A4
reference
has
failed.
Check the input to the
A5
synthesizer from the
A4
reference.
200
kHz reference missing
at
synth
The
A5
module indicates that the
200
kHz reference
signal from the
A4
module is not being detected.
Either the
A4
module has failed to output the
200
kHz reference signal, or the
A5
module
is
failing to
detect the signal.
Check the
200
kHz reference output of the
A4
module
at
the input to the
A5
module.
Reference out-of -lock at
10
MHz
The
A4
module indicates that the
10
MHz reference
VCO
is out-of-lock.
Either the
A4
module has failed or a bad external
reference is connected.
Disconnect any external reference and power-up the
instrument again or replace the
A4
module
if
the
error is still present with no external reference.
Service
Error
Messages
50-5