IC
Operation
Messages
This chapter provides descriptions for both front panel and HP-IB
operation messages. (For information about service messages,
numbered
500
and above, refer to Chapter
5c,
"Service Error
Messages.
Front Panel
Operation
Messages
SEQ
X
SAVE
- -
XXX
registers available
This message is displayed when the
inform you of how many registers are still available.
If
a register is available, enter the two-digit number of the
register you wish to save.
key is pressed to
SEq
XREG--XXhas not beensaved
This message is displayed when an attempt is made to recall
a register that has not been saved in the sequence. Check to
be sure that the appropriate sequence is selected and that
you have entered the correct register number.
SEQ
XDEL--Enternumberto delete
This message is displayed when the
(DEL)
key is pressed.
Enter the number of the memory register you wish to
delete. When a register is deleted, the settings saved in it
are erased and the register number is removed from the
sequence.
SEQXhasnoregisters savedinit
This message is displayed when a sequence is selected that
has no registers saved in it.
If
you wish to save registers in
the sequence, set-up the instrument, press the
ISAV)
key, and
enter a two-digit register number.
001
No
externaldc coupling
for
PM
This message is displayed when PM
is
selected and
(EXTDC)
or
(1
kHz
+
EXT
DC)
is also selected.
DC
coupling
of an external source is not possible for PM.
If
you press
Or, select
[EXT]
coupling for PM. Additional internal plus
external modulation capabilities (such
as
[I
kHz
+
EXT
AC))
are available through HP-IB control
of
the instrument; refer
to Chapter
2,
"HP-IB Programming.
(1
kHz
+
EXT
DC]
you
will
actually get
1
kHz and external
ac.
Operation Messages
1c.l