002
Modulation exceeds deviationrange
This message is displayed when modulation is set to a level
that exceeds the operating range
of
the instrument. This
condition occurs when a modulation level is entered that is
out-of-range for the current RF frequency setting,
or
when
the RF frequency setting is changed and the modulation
setting is out-of-range for the new setting.
003
There areno registers available
This message is displayed when an attempt
is
made to save
a memory register and all of the instrument’s memory
registers have already been used. Delete
any
unneeded
registers in order to save new ones. Deleting registers from
any sequence
will
make them available for saving new
settings in the sequence you are using.
004
Invalid units selection
This message is displayed when a units key is pressed that is
not valid for the active function. Check that the units key
you select is labeled with the appropriate units for the value
you are entering.
Increment value entry out of range
This message is displayed when the
LSET)
key is pressed
and a value is entered that is not within the increment
value range for the active function. Refer to “Increment
Set” in Chapter lb, “Operation Reference,” for a listing
of
the increment value ranges.
005
006
End of increment range
This message is displayed when the
arrow key is pressed and the increment value does not set
the instrument to a setting that
is
within the instrument’s
allowable range. To view
or
change the increment value,
press the
(-SET_)
key. Refer to “Increment Set’’ in
Chapter
lb,
“Operation Reference, for a listing of the
increment value ranges.
or
@)
increment
007
Enteredvalueoutof range
This message is displayed when a value is entered that
does not set the RF output signal within the instrument’s
allowable range. Refer to “Function” in Chapter
lb,
“Operation Reference,
for information on the instrument’s
allowable ranges.
008
Amplitude exceeds specified range
This message is displayed when the instrument’s amplitude
is set to a level that exceeds
+
10 dBm.
010
End
of
knob range
This message is displayed when the knob is turned but
changing the selected digit would set the instrument to a
value that is not within its allowable range.
lc.2
Operation Messages