CW
Frequency
Accuracy
Performance
Tkst
(Option
1E5
Only)
Connect the Test
Equipment
HP
8647A
under
test
Measuring
Receiver
Figure
8-12.
CW
Frequency Accuracy Equipment Setup
Configure the
1.
Reset
Measuring Receiver
2.
Frequency mode
3.
Special Function
7.1.
Press
(7J
0
Configure the
HP
1.
Turn modulation
off;
press
(MOD
ON~OFF).
8647A
2.
Set the amplitude; press
@
m.
3.
Turn the RF output
on;
press
Cm).
Measure the
Frequency
1.
Enter the frequencies
shown
in the test record.
2.
Record the test results and compare the results to the limits in the
test record.