Memory Interface
2.
Copy
Arrow
Keys
Press these keys
to
initiate
a
copy
of
the memory registers between
the instrument and the HP 83301A. After the instrument has
successfully completed the copy, it will display a message to let you
know.
Making
a
Copy
When the HP 83301A is connected to the instrument, press the
@)
to
copy all of the memory registers saved in the instrument into the
HP 83301A's memory. After you have pressed
an
arrow key, you must
also press the
(SAV)
key
on
the instrument's front panel to begin the
copy process. Press the
to
copy the memory registers stored in the
HP 83301A into the instrument's memory.
Note
Copying memory into the instrument
or
the HP 83301A causes any
existing memory registers in the receiving device to be erased. It does
not effect the memory in the sending device, however.
3.
BUSY
This light is turned on while memory registers are being copied from
one device
to
another. Attempting to operate the instrument or
memory interface while this light is on may cause the memory data to
be corrupted. The light will turn
off
when the copy is complete.
If
the Busy light flashes for
a
few seconds and then turns off, the
instrument did not find valid memory register data in the memory
interface. If the Busy light continues to
flash,
or
if
it does not turn on
at all after an arrow key and the
ISAV)
key have been pressed, refer to
Chapter 5b, "Troubleshooting Information.