Table 2 continued. . .
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment –
Guidance
Conducted
RF IEC
61000-4-6
Radiated
RF IEC
61000-4-3
3Vrms 150 kHz
to 80 MHz
3 V/m 80 MHz
to 2.5 GHz
3Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the iCR 3600, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separations
distance
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800 MHz
d = 2.3
√
P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters as determined by an
electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflections from structures, objects
and people.
continued on next page. . .