Ecotec E3000 Messages 6-3
kina22e chapter 6.fm technical handbook(1206)
E25 Remove sniffer from CAL port Sniffer inserted in calibration port during
start-up
or
sniffer needs to be removed during
internal calibration process
Remove sniffer from calibration port
of ECO-Check!
Light barrier of ECO-Check dirty Blow out with fresh air and / or clean
with cotton swab!
W28 Real time clock reset! Please enter
date and time!
CPU-board has been replaced Please enter date and time!
See Section 4.4.5
Battery on CPU-board faulty Replace CPU-board*
E30 Sensitivity too low Argon signal too low Check if fresh air is provided to
sniffer line
Please recalibrate the Ecotec E3000
W31 Factor K1 out of range! Interfering gases other than
cyclopentane or isopentane (e.g.
alcohols) detected
Recalibrate IGS! (see Section 8.2)
W34 Flow has changed! Flow has changed by more than 30%
since last calibration (warning will
disappear if change drops below 20%
again).
Recalibrate the
Ecotec E3000
(See Section 3.5) or replace sniffer
line filters!
W35 Flow through capillary is too low Actual flow is lower than lower flow limits
(In
LOW FLOW mode)
Filter in sniffer line blocked
Capillary blocked
Filter in main unit blocked
Lower flow limit setting too high
Replace sniffer tip filter
See Section 8.6
Replace filters of sniffer line and
recalibrate! (felt filter, capillary filter
and / or filter pad in SL3000XL
sniffer line and recalibrate) See
Section 8.6
or
Replace probe cable*
Replace internal filter*
Decrease lower flow limit
See Section 4.4.1
E36 Capillary broken (>2 s) Actual flow is higher than upper flow
limits for more than 2 s
Capillary broken or leaky
Upper pressure limit setting too low
Replace sniffer line and recalibrate!
Or: Replace probe cable* and
recalibrate!
Increase upper flow limit (see
Section 4.5.1
Press OK to restart the Ecotec
E3000
E37 Capillary broken (>10 s) Actual flow is higher than upper flow
limits for more than 10 s, emission will be
switched off to protect filaments
Capillary broken or leaky
Upper pressure limit setting too low
Replace sniffer line and recalibrate!
Or: Replace probe cable* and
recalibrate!
Increase upper flow limit (see
Section 4.5.1)
Error no. Plain text message Possible reason Recommendation