6-4 Ecotec E3000 Messages
kina22e chapter 6.fm technical handbook(1206)
E38 Capillary broken (>60 s) Actual flow is higher than upper flow
limits for more than 60 s, pumps will be
switched off to protect filaments
Capillary broken or leaky
Upper pressure limit setting too low
Replace sniffer line and recalibrate!
Or: Replace probe cable* and
recalibrate!
Increase upper flow limit (see
Section 4.5.1)
Press OK to restart the Ecotec
E3000
E39 No emission! Start emission failed on both filaments
during start-up!
After extended shutsown times this
error may occur during the first
10 min after power on
Please acknowledge the error and
startup again automatically.
If problem persists, call nearest
INFICON service representative.
E40 No emission! Emission failed during operation After extended shutdown times this
error may occur during the first 10
min after power on.
Please acknowledge the error and
startup again automatically.
If problem persists, call nearest
INFICON service representative.
E41 No communication with Transpector! Software cannot establish
communication with Transpector
Check cable connection between
Transpector and CPU-board!*
Call nearest INFICON service
representative!
E42 Transpector temperature > 70°C or <
0°C!
Air filter dirty
Ambient temperature too high
Ambient temperature too low
Replace main air filter (see Section
8.3.)
Check ambient temperature
E43 Transpector limit exceeded! Refer to Transpector internal data Call nearest INFICON service
representative!
E44 Transpector communication error! Refer to Transpector internal data Call nearest INFICON service
representative!
E45 Transpector hardware error! Refer to Transpector internal data Call nearest INFICON service
representative!
E46 Transpector hardware warning! Refer to Transpector internal data Call nearest INFICON service
representative!
E47 Transpector over-pressure! After extended shutdown times this error
may occur during the first 10 min after
power on.
This may also occur if sniffer line has
been disconnected for extended times!
Please acknowledge the error and
the Ecotec E3000 will startup again
automatically.
Reconnect sniffer line and
acknowledge error, Ecotec E3000
will startup again automatically.
E48 Transpector no emission! After extended shutdown times this error
may occur during the first 10 min after
power on.
This may also occur if sniffer line has
been disconnected for extended times!
Please acknowledge the error and
the Ecotec E3000 will startup again
automatically.
Reconnect sniffer line and
acknowledge error, Ecotec 3000 will
startup again automatically.
Error no. Plain text message Possible reason Recommendation