Ecotec E3000 Messages 6-5
kina22e chapter 6.fm technical handbook(1206)
W49 No emission with first filament! Start emission failed with first filament ! Switch to 4.8
E50 Unkown TC-Error Failure of the turbo electronic drive Call nearest INFICON service
representative!
E51 No communication with turbo controller Software cannot establish
communication with controller of TMP
Call nearest INFICON service
representative!
W59 EEPROM parameter queue overflow! May occur if software “update” to older
version is performed
Restart the Ecotec 3000, if problem
persists, call nearest INFICON
service representative!
W60 All EEPROM parameter lost! Please
check your settings!
New EEPROM has been installed,
EEPROM on motherboard is virgin
All settings in software menu are
reset to default! Please enter your
settings again!
If message comes up repeatedly during
start-up, EEPROM on motherboard is
faulty
Replace EEPROM*
W61 EEPROM parameter initialized! Software update performed and new
parameters have been introduced
Newly introduced parameters are listed
below warning
Acknowledge warning
If message comes up repeatedly during
start-up, EEPROM on motherboard is
faulty
Replace EEPROM*
W62 EEPROM parameter lost! Parameter has been modified during
software update and reset to default
value
Affected parameters are listed below
warning
Check setting of modified
parameters in corresponding
software menu and set to desired
value!
If message comes up repeatedly during
start-up, EEPROM on motherboard is
faulty
Replace EEPROM*
W63 TSP parameter mismatch! TSP replaced or SW updated
Non volatile memory of TSP faulty
EEPROM on motherboard faulty
Please check data and restore!*
Call nearest INFICON service
representative!
W64 There are outstanding warnings! Acknowledged but still valid warnings will
be repeated every 2 hrs or on each new
power on
Please double-check the warnings!
W66 ECO-Check virgin! New ECO-Check reference leak installed
in main unit
Please enter serial number and
code.
See Section 4.7, Replace ECO-
Check
W67 ECO-Check will be expired at
DD.MM.YYYY
Warning for pending expiration will be
issued starting three months before
actual expiration
Please order new ECO-Check gas
reservoir! (cat. no. 521-010)
W68 ECO-Check expired! ECO-Check has been used for more
than 1 year or has been manufactured
more than 2 years ago.
Please replace ECO-Check gas
reservoir!
See Section 8.8
W70 All EEPROM parameter of ECO-Check
lost!
EEPROM in ECO-Check is empty or
faulty
Replace ECO-Check!
Error no. Plain text message Possible reason Recommendation