Maintenance 8-16
kina22e chapter 8.fm technical handbook(1206)
Notice: Screw in the new reservoir with hex nut key only!
4 Re-install the ECO-Check in the Ecotec E3000.
Notice: The ECO-Check does not fit in the Ecotec E3000 completely. There is a
little gap between the front panel of the Ecotec E3000 and the ECO-Check.
On the certificate which is delivered with the replacement gas reservoir you will find
a new serial number and a 12-digit-code containing the new leak rate and other leak
specific information. In the software menu please go to
HISTORY & MAINTENANCE /
REPLACE ECO-CHECK. In the opening sub-menu phase enter the new serial number
in the first line and the 12-digit-code in the second line and press okay.
Fig. 8-18 Installing the gas reservoir
Fig. 8-19 Gap between ECO-Check and front panel