General Information 1-17
kina22e chapter 1.fm technical handbook(1206)
Water protection tip (optional)
If you intend to perform leak testing on parts that are not completely dry (e.g. due to
condensation after performance testing), we strongly recommend to use a water
protection tip.
To install the water protection tip,
1 screw off the metallic capillary filter at the very end of the sniffer tip and
2 install the water protection tip instead.
Notice: Please do not forget to re-install the little rubber seal when switching to the
water protection tip.
ECO-Check Reference leak (optional),
for helium and hydrogen only
If detecting helium or hydrogen, a ECO-Check reference leak may be used for
internal calibration. The ECO-Check reference leak has to be inserted to the Ecotec
E3000 (see section 2.3.3).
On first usage of your ECO-Check you need to initialize the use of this reference leak
in the Ecotec E3000 menu.
Please perform the following steps:
1 Insert the ECO-Check into the appropriate opening of the Ecotec E3000
2 In the software menu go to HISTORY & MAINTENANCE /REPLACE ECO-CHECK.
Notice: The Ecotec E3000 must be set to
USER MODEADVANCED“ for access to this
submenu (See section 4.5.3). Go to
SETTINGS / DISPLAY / USER MODE.
3 On the certificate, which is delivered with the ECO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the second line and press OK.
Fig. 1-7 Installing water protection tip