MEDIAJET – Operating instructions V09 Operation
25
www.integra-biosciences.com
4 Operation
Purpose This chapter describes how to adapt the general device settings and how to
program and carry out a petri dish filling process.
4.1 Overview MAIN MENU functions
The list below provides an overview of the selectable functions in the MAIN
MENU of the operating panel.
Press >>> in order to reach the second page of the MAIN MENU.
FILL DISHES: To adapt and store (see “4.4
Adapting a dish filling program” on
page 30), as well as execute (see “4.5 Run dish filling programon page 34 and
“4.9
Mixing additives into the agaron page 42) up to 19 individual Petri dish
filling programs.
LOAD CAROUSEL: This function supports the operator to load the carousel
with empty Petri dishes (see4.3
Loading Petri dishes” on page 28) and remove
the filled agar plates after termination of the dispensing process (see “4.6
Removing filled Petri dishes” on page 38).
TUBE FILLER: Requires the optional available accessory TUBEFILLER, which
allows to automatically fill test-tube racks on MEDIAJET.
This function allows the operator to adapt, store and execute 19 individual test-
tube filling programs (see “4.10
TUBEFILLER option on page 44 and “7.8
TUBEFILLER option” on page 64 for more information).
CLEANING: The piston of the feeder or stacker can be moved for cleaning by
this function (see “5.1
Cleaning” on page 48).
DOSING FUNCTION: Function for independent volumetric dosing applications
using the peristaltic pump of MEDIAJET (see “4.8
Dosing functionon page 41).
SYSTEM PARAMETER: To adapt the general device parameters (see “4.2
Adapt system parameter” on page 26).

Related product manuals