MEDIAJET – Operating instructions V09 Operation
26 INTEGRA Biosciences AG
4.2 Adapt system parameter
The following table gives an overview about the other parameters to be set
under the menu option SYSTEM PARAMETER, if required.
System
parameter
Description and function
Language
selection
Sets the desired language.
Time/date Sets the local time and date.
UV lamp Displays the operating hours of the UV-lamp.
Allows to switch the UV-lamp on or off and to reset the
operating-hour counter.
Display
contrast
Altering the display contrast.
Protocol
printing
Provides choice of printing the process data on an independent
external printer or on MEDIACLAVE and to execute a test print
(see more in “4.7
Process documentation” on page 39).
Program
configura-
tion
Enabling/disabling functions in program parameters:
• Turbo mode: OFF
• Shaker level (Pour Plate): OFFBatch mode: OFF
• Power Stacker: OFF
(Allows handling stack weights that are at the upper limit of
the stacker force, e.g. filling Petri dishes with large volumes
of agar medium.)
• Pump Options: OFF
(Enabling this function allows pumping back the medium to
the source.)
• Biplate alignment (only MEDIAJET vario with biplate
option): OFF
Cooler
(optional)
Allows to switch off the cooling plate, in case the option is
installed.
Dish
sensor
teach-in
(optional)
Procedure to adjust the optical dish sensor, see “4.2.2
Dish
sensor teach-in (teach-in sensors only)” on page 27.
Biplate
(optional)
To fill Petri dishes with two compartments, see “4.2.3
Biplate
(optional)” on page 27.
Info General information on the device:
• Software and electronics version
• Series number
• Operating-hour and dish counter.
System
messages
Displays information on the status of the device (only shown, if
a message available).