MEDIAJET – Operating instructions V09 Installation
21
www.integra-biosciences.com
3.4.5 Base plate for filling nozzle holder
For filling of Petri dishes with a height above 21 mm a base plate, used as a
spacer, must be mounted.
For MEDIAJET vario with biplate option, a special base plate must be mounted,
if Petri dishes with a height above 21 mm are filled:
3.5 Biplate option
The biplate option allows the filling of two Petri dish compartments in parallel by
two filling nozzles and a double pump head (available only with MEDIAJET vario
with biplate option, Part No. 113 002).
Remove the two screws below the caps on
the filling nozzle holder.
Place the base plate under the filling nozzle
holder and fix it with the long screws
supplied with the base plate.
Put the screw caps on the screws.
Remove the two screws of the filling nozzle
holder.
Place the base plate for biplate option under
the filling nozzle holder, so that the two L-
shaped openings of the base plate are
stacked below those of the filling nozzle
holder.
Fix the stack with the screws of the filling
nozzle holder.
The filling nozzle holder must be adjusted,
see “4.2.3
Biplate (optional)” on page 27.
Enable the biplate alignment under program
configuration, see “4.2 Adapt system
parameter” on page 26.