MEDIAJET – Operating instructions V09 Operation
27
www.integra-biosciences.com
4.2.1 Language selection
Press >>> in order to reach the second page of the MAIN MENU.
Select a language under SYSTEM PARAMETER / LANGUAGE
SELECTION.
4.2.2 Dish sensor teach-in (teach-in sensors only)
The dish sensor teach-in is a procedure to adjust the optical sensor to the used
type of Petri dishes. Whenever the type of Petri dishes is replaced (different
brand or size) the teach-in procedure should be carried out.
Select DISH SENSOR TEACH-IN under SYSTEM PARAMETER.
4.2.3 Biplate (optional)
This function supports the adjustment of the filling nozzle holder for dispensing
into two-compartment Petri dishes and tests the automatic alignment of the
dividing wall of the dishes. Before you perform a biplate filling program, follow
this procedure:
Select PROGRAM CONFIGURATION under SYSTEM PARAMETER and
enable the biplate alignment.
Press SAVE.
The screen will appear in the language
selected.
Follow the instructions on the screen. Press
START and later CONTINUE.
If “Teach-in successful” is displayed press
END.
Select BIPLATE under SYSTEM
PARAMETER.
Initialise the rotor by pressing INIT ROTOR.