MEDIAJET – Operating instructions V09 Maintenance
49
www.integra-biosciences.com
Remove the O-ring of the drive wheel by dismounting the filling nozzle holder
and wash it with water. Replace the O-ring if, if necessary (only MEDIAJET
vario with biplate option).
Subsequently disinfect the filling chamber using 70 % Ethanol.
Regularly apply a very thin coat of grease (Klüber Microlube GBU-Y131, Part
No. 200 150) to the anchor bolts of the rotor/carousel axis.
If necessary, the CLEANING dialogue also allows you to clean beneath the
feeder and stacker pistons.
The MEDIAJET is now clean and ready for another filling process.
The moving parts of the pumphead should
be checked from time to time for freedom of
movement.
Occasionally lightly grease the lever (1) and
the rollers with Teflon lubrication oil (see
image).
Keep STACKER or FEEDER pressed and
move the corresponding piston using the
ARROW KEYS.
Clean the lifter rod of feeder and stacker
(the rear side of the dish plate in particular).
Apply a very thin coat of grease (Klüber
Microlube GBU-Y131, Part No. 200 150) to
the feeder/stacker pistons above the
sealing. Place the feeder and stacker piston
back to the starting position.
Re-install rotor, safety cover and carousel.