VisualEyes515/525 Instructions for Use - US Page 41
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ORION is intended for use in the electromagnetic environment specified below. The customer or the
user of the ORION should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Compliance
Electromagnetic environment -
guidance
Electrostatic
Discharge (ESD)
IEC 61000
-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be greater than
30%.
Electrical fast
transient/burst
IEC61000
-4-4
± 2 kV for power supply
lines
100 kHz repetition
frequency
± 1 kV Line-to-line
100 kHz repetition
frequency
± 2 kV
± 1 kV
Mains power quality should be that
of a typical
commercial or residential
environment.
Surge
IEC 61000-4-5
± 1 kV Line-to-line
± 2 kV Line-to-ground
± 1 kV
± 2 kV
Mains power quality should be that
of a typical commercial or residential
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000
-4-11
0% UT for 0.5 cycle
0 % UT for 1 cycle
and
70% UT
for 25/30 cycles
Single phase: at 0°
0% UT for 0.5 cycle
0 % UT for 1 cycle
and
70% UT
for 25/30 cycles
Single phase: at 0°
Mains power quality should be that
of a typical commercial or residential
environment. If the user of the
ORION
requires continued operation
during power mains interruptions, it
is recommended that the ORION be
powered from an uninterruptable
power supply or its battery.
Power frequency
(50/60 Hz)
IEC 61000
-4-8
30 A/m 30 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or residential
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.