EasyManua.ls Logo

Interacoustics VisualEyes 515 - Page 55

Interacoustics VisualEyes 515
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-0126440-A – 2020/08
VisualEyes515/525 Instructions for Use - US Page 42
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The ORION is intended for use in the electromagnetic environment specified below. The customer or the
user of the ORION should assure that it is used in such an environment,
Immunity test
IEC / EN 60601 test
level
Compliance level
Electromagnetic environment
guidance
Portable and mobile RF
communications equipment should
be used no closer to any parts of the
ORION, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance:
Conducted RF
3 Vrms
3 Vrms
IEC / EN 61000-4-6
150kHz to 80 MHz
6 Vrms in ISM bands
150kHz to 80 MHz
80 % AM at 1 kHz
6 Vrms
Radiated RF
3 V/m
3 V/m
80 MHz to 800 MHz
IEC / EN 61000-4-3
80 MHz to 2,7 GHz
80 % AM at 1 kHz
800 MHz to 2,7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies

Table of Contents