PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 103 of 141
88292-EN, Rev D
LIO Treatment Procedure
Perform the following procedure in conjunction with the instructions in the Laser Indirect
Ophthalmoscope (LIO) operator manual.
1.
Select the Port where the LIO is connected.
2.
Select READY mode. The aiming beam turns on.
3.
Select the aiming beam intensity.
4.
Select the appropriate exposure time.
5.
Select the appropriate laser power.
6.
Make any necessary adjustments to the headset to ensure safe and clear view of the
retina, as described in the LIO operator manual.
7.
Position the aspheric lens and select the spot diameter, as described in the LIO operator
manual.
NOTE
When using the LIO, the laser spot diameter cannot be adjusted. Spot diameter is
determined by the choice of aspheric lens and the position of the LIO relative to the
target tissue. See the LIO operator manual for detailed information.
8.
Depress the footswitch to deliver the treatment laser beam to the tissue.