PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 122 of 141
88292-EN, Rev D
System Relocation Instructions
To move the system to another location:
1.
Ensure that the wall circuit breaker is turned off and the key switch is in the OFF position.
2.
Remove the power cable from the wall outlet and the system main power receptacle.
3.
If a remote door interlock is utilized, remove the interlock plug and cable from the interlock
port and transport separately.
4.
Disconnect the footswitch cable from the footswitch port and transport separately. Never
drag the footswitch.
5.
Position the console a minimum of 15 cm (» 6 in) away from walls, furniture or other
equipment. Adequate space around the console ensures proper air circulation for system
cooling purposes.
6.
If there are changes in environmental conditions (temperature or humidity) allow system
to acclimate for 4 hours prior to use.
Room Preparation
1.
Verify that the system power cable and plug are correctly connected, as instructed in the
preoperative instructions.
2.
Verify environmental conditions are within limits for operation.
For systems configured with an electrical wall circuit breaker, always place the electrical wall
circuit breaker in the off position before inserting the plug into the receptacle.
3.
Verify that the electrical service is turned on.
4.
Verify that the laser warning sign has been posted outside of the treatment room door.
5.
Ensure that all attending personnel in the treatment room are wearing
appropriate eye protection goggles or eyeglasses.