PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 104 of 141
88292-EN, Rev D
Between Patient Treatments
At the completion of each patient treatment:
1.
Press <End Treatment>to exit the Treatment screen. If the treatment report feature is
enabled, the Treatment Report window displays. Otherwise, the Home screen displays.
NOTE
If no laser shots were fired during the treatment session, pressing <End
Treatment>returns the system to the Home screen whether or not the treatment report
feature is enabled.
2.
Disinfect the chin rest and head rest using mild soap and water. Dry with a soft cloth.
3.
Disinfect the contact lens following the contact lens manufacturer instructions.
System Shut-down
At the end of the day, or during an extended period of inactivity:
1.
Shut down the system as described in “Shutting Down the System”.
2.
Remove the key to prevent unauthorized use of the system.
3.
Clean the system as described in “User Maintenance”.
4.
Place a dust cover over the slit lamp.