Electromagnetic Compatibility
Applicable to PASCAL Synthesis Laser Systems with serial numbers starting with “80” (eg
80xxxxxx).
This product conforms to the EMC standard (IEC 60601-1-2: 2014). The expected
electromagnetic environment for the whole life cycle is professional healthcare facility
environment.
a)
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided
in the ACCOMPANYING DOCUMENTS.
b)
Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
c)
The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM
should be observed to verify normal operation in the configuration in which it will be
used.
d)
The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS
other than those specified may result in increased EMISSION or decreased
IMMUNITY of this EQUIPMENT and result in improper operation.
e)
Do not use the devices generating electromagnetic waves within 30cm from all the
parts of the instrument and system. Those devices may have influence on this
product.
HEAD UP DISPLAY SYSTEM HUD-1 Module
PASCAL Laser Indirect Ophthalmoscope (LIO)
AC power cord (for SLIT LAMP)
AC power cord (for LASER CONSOLE)
SIP/SOP Cable (for Footswitch)