PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 111 of 141
88292-EN, Rev D
System does not turn on.
System is not plugged in.
Plug in system. Verify power cable is well seated in wall outlet
and in main power receptacle.
Wall circuit breaker is in off position.
Turn on wall circuit breaker.
Key is absent or in OFF position.
Insert key and rotate to ON position.
Turn key switch to OFF position, wait at least one minute and
then turn to ON position. If system fails to start, contact
service.
System monitor is blank for more than 30 seconds.
Internal system error during boot-up.
Turn key switch to OFF position, wait at least one minute and
then turn to ON position.
Monitor/touchscreen panel power, USB and signal cables are
not correctly connected.
Ensure that monitor cables are correctly connected.