EasyManua.ls Logo

Iridex PASCAL Synthesis - Instructions for Use; Intended Use Environment; Patient Environment

Default Icon
141 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 123 of 141
88292-EN, Rev D
Instructions for Use
Intended Use Environment
The intended Use Environment of the laser system is as follows:
Professional Healthcare Facility Environment
Physician offices, clinics, multiple treatment facilities, hospitals except for near active
HF surgical equipment and the RF shielded room of an ME system for magnetic
resonance imaging, where the intensity of EM disturbances is high.
Patient Environment
The Patient Environment is defined by the following:
When the patient or inspector comes in contact with the devices (including the
connecting devices)
When the patient or inspector touches the person that comes into contact with the
devices (including the connecting devices).
In the patient environment (shown below) , use the device conforming to IEC
60601-1 standard. If you are compelled to use any device not conforming to the
standard, use an insulation transformer which conforms to IEC 60601-1
standard.

Table of Contents

Related product manuals