PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 105 of 141
88292-EN, Rev D
Maintenance Instructions
To assure that the system remains safe with regard to electromagnetic disturbance
throughout the expected service life, the following actions are recommended.
Annual Maintenance
Preventive maintenance, safety, power and calibration checks should be performed annually
by Iridex Corporation-certified personnel to ensure proper laser performance.
System Repair
All repairs should be performed by certified personnel to ensure proper system performance.
User Maintenance
The following maintenance procedures are to be performed by the user to ensure proper
performance of the system.
Cleaning the Console External Surfaces
Clean the external surfaces of the console daily, after use. Use a cloth dampened with a non-
caustic cleaning solution (e.g., soap and water) to clean the external non-optical surfaces of
the console. Dry with a clean cloth or allow to air dry. Do not spray or pour cleaning agents
directly onto the console.
Cleaning the Control Panel Screen
Use a soft, dry cloth to apply antistatic glass or plastic cleaner to the control panel screen.
Maintaining Effectiveness of Grounds
Clean the unplugged power cord to maintain the protective earth. Use a soft, dry cloth.