PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 23 of 141
88292-EN, Rev D
It may take longer for the equipment to achieve a ready state in a low-temperature
environment.
It is the responsibility of the physician to verify that the aiming beam spot visualized through
the slit lamp is the expected size. If the aiming beam size or pattern appears inappropriate
or distorted, do not proceed with treatment. Readjust the slit lamp focus. If the problem
persists, contact service.
It is the responsibility of the physician to select the appropriate power and treatment
location. The lowest practical setting should always be used to achieve the desired clinical
outcome.
Do not use a wet cloth to clean the control panel screen. Doing so may damage the screen.